BLOOD BANK REFRIGERATOR

The Osworld Blood Bank Refrigerator is designed for the precise requirement of Blood Banking and processing applications.

Features
  • Temp. Control: Microprocessor-based PID Control with Auto-tune, CE Marked
  • Temp. Sensor: PT-100 RTD, Class "A"
  • Temp. Display: Digital LED, 3½ digit
  • Heating: U-shaped Nichrome wire in SS Sheathing
  • Cooling: CFC Free compressor utilizing R134-a eco friendly refrigerant, with condenser, motor and relay. Complete Unit – Copeland Make.
  • Air Circulation: Flange motor with Impeller/ Blower
  • Chamber Illumination: Fluorescent light with door switch
  • Feet: Castor Wheels
  • Trays: SS Wire Mesh, Heavy Duty
  • Construction: Double-walled with Poly Urethene (PUF) insulation provided with outer stainless steel metal door and inner glass viewing door.
Specifications
  OBR S/G-4 OBR S/G-8 OBR S/G-12 OBR S/G-16 OBR S/G-21 OBR S/G-28 OBR S/G-34
Temp. Range 2°C to 8°C
Temp. Resolution ±0.1°C ±0.1°C ±0.1°C ±0.1°C ±0.1°C ±0.1°C ±0.1°C
Temp. Accuracy ±0.2°C ±0.2°C ±0.2°C ±0.2°C ±0.2°C ±0.2°C ±0.2°C
Temp. Uniformity ±1°C ±1°C ±1°C ±1°C ±1°C ±1°C ±0.7°C
Internal Size (W x D x H) cm 45 x 45 x60 60 x 60 x60 60 x 60 x 90 60 x 60 x125 60 x 80 x125 80 x 80 x 125 80 x 80 x155
External Size (W x D x H) cm 65 x 114 x 130 79 x 121 x 130 79 x 121 x 165 79 x 121 x 199 76 x 141 x 199 96 x 141 x 199 96 x 141 x 225
Capacity (Litres) 120 200 324 450 600 800 1000
Shelves (S.S.) 2 2 3 4 5 5 5
Shipping Weight - kg (lb) 200 (440) 240 (528) 250 (550) 265 (583) 295 (649) 325 (715) 375 (825)
Input Voltage 230 Volts AC, 50 Hz
Material Inner Chamber: SS 304 Grade/ Mirror finish
  Outer Chamber: Mild Steel Powder Coated/ SS 304 Grade

Installation Mandatory Requirement
  • Room temperature around machine should preferably be at 25°C with air conditioning or a well ventilated room with an exhaust fan. However, the surrounding temperature should not exceed 30°C.
  • Stabilized Input Voltage of 230V AC, 20 Amps. Use of a Servo Controlled Stabilizer is recommended.
Calibration / Validation
i) Calibration
  • Temperature sensor probe calibration with traceability to ERTL
  • Temperature controller calibration with traceability to ERTL
ii) Validation
  • IQ, OQ and PQ documentation and protocols shall be provided with equipment.
  • Documentation prepared / developed for Osworld by Premier Validation Ltd. – Ireland, one of Europe’s leading validation consultants
  • Performance Qualification with Temperature Mapping on site. The Performance Validation Test consists of one cycle at any one temperature point for an 8-hour period. Data is collected in printed format from up to six different points within the chamber. Temperature uniformity is considered acceptable if the deviation is less than ±1°C of set temperature.
Optional Extras / Accessories :
  • Oslog Data Storage Device and Oslog 21 CFR Part 11 Compliant Software - Both have been developed exclusively for Osworld by M/s Godrej & Boyce Mfg. Co. Ltd.
  • Additional high-temperature safety system to cut off all heaters with audio-visual alarm
  • PLC-based system: PLC with HMI 3.7” touch screen grey monitor with 21 CFR Part 11 features (OSLOG) software.
  • Door access security system
  • Voltage Stabilizer for above unit
  • GSM System: Alarms sent to predefined mobile numbers
  • HMI: 5.7" touch screen grey in lieu of 3.7"
  • Controller with printer interface to connect to EPSON Dot Matrix line printer. Date, time and temperature can be printed whereas print interval can be set between 1 to 99 minutes
  • Data Logger: 4 point temp. Data logger with sensors placed at specific points in chamber with printer interface and memory
Download Brochure