DEEP FREEZER

The Osworld Deep Freezer is suitable for medical and scientific applications (ex. Reagents, biologicals, pharmaceuticals and other commonly used laboratory materials). These Freezers provide energy efficient, convenient, safe and reliable performance for optimal storage temperature environments necessary for a wide range of life science, pharmacy, biological, medical, clinical and industrial applications.


walk In stability chamber manufacturer in india
Features
  • Temp. Control: Microprocessor based PID Control with Auto-tune, CE Marked
  • Temp. Sensor: PT-100 RTD, Class A
  • Temp. Display: Digital LED, 3½ digit
  • Cooling: CFC Free Copeland Make compressor utilizing R134-a eco friendly refrigerant with condenser motor and relay.
  • Air Circulation – Flange Motor with Impeller/ Blower
  • Feet: Castor Wheels
  • Trays: SS Wire Mesh, heavy duty
  • Construction: Double-walled with PUF insulation. Outer door lockable.
  ODF - 45
Temperature Range Ambient to -20°C
Temp. Accuracy ±0.5°C ±0.5°C ±0.5°C ±0.5°C ±0.5°C
Temp. Uniformity ±1°C ±1°C ±1°C ±1°C ±1°C
Temp. Resolution +0.1°C +0.1°C +0.1°C +0.1°C +0.1°C
Internal Size (W x D x H) cm 45 x 45 x 60 60 x 60 x 60 60 x 60 x 90 60 x 60 x 125 60 x 80 x 125
External Size (W x D x H) cm 65 x 114 x 130 79 x 121 x 130 79 x 121 x 165 79 x 121 x199 76 x 141 x 199
Capacity (Litres) 120 200 324 450 600
Shelves (S.S.) 2 2 3 4 5
Input Voltage 230 Volts AC, 50 Hz
Material Inner Chamber: SS 304 Grade/ Mirror finish
  Outer Chamber: Mild Steel Powder Coated/ SS 304 Grade

 

Pre-Installation Mandatory Requirement
  1. Room temperature around machine should preferably be at 25°C with air conditioning or a well ventilated room with an exhaust fan.However, the surrounding temperature should not exceed 30°C.
  2. Stabilized Input Voltage of 230V AC, 20 Amps. Use of a Servo Controlled Stabilizer is recommended.
Calibration / Validation
i) Calibration
  • Temperature sensor probe calibration with traceability to ERTL
  • Temperature controller calibration with traceability to ERTL
ii) Validation
  • IQ, OQ and PQ documentation and protocols shall be provided with equipment.
  • Documentation prepared / developed for Osworld by Premier Validation Ltd. – Ireland, one of Europe’s leading validation consultants
  • Performance Qualification with Temperature Mapping on site. The Performance Validation Test consists of one cycle at any one temperature point for an 8-hour period. Data is collected in printed format from up to six different points within the chamber. Temperature uniformity is considered acceptable if the deviation is less than ±1°C of set temperature.
Optional Extras/Accessories
  • Oslog Data Storage Device and Oslog 21 CFR Part 11 Compliant Software - Both have been developed exclusively for Osworld by M/s Godrej & Boyce Mfg. Co. Ltd.
  • Additional high-temperature safety system to cut off all heaters with audio-visual alarm
  • Voltage Stabilizer for above unit
  • Data Logger: 4 point temp. Data logger with sensors placed at specific points in chamber with printer interface and memory
Download Brochure