WALK IN STABILITY CHAMBER

The Osworld Walk in Stability Chamber is basically designed for testing of production samples where control of temperature and relative humidity are desired.

Features
  • Temp. Control: Microprocessor-based PID Control with Auto-tune, CE Marked
  • Temp. Sensor: PT-100 RTD, Class "A"
  • Temp. Display: Digital LED, 3½ digit
  • Heating: U-shaped Nichrome wire heater in SS Sheathing
  • Cooling: CFC Free compressor utilizing R134-a eco friendly refrigerant, with condenser, motor and relay. Complete Unit – Copeland Make.
  • Air Circulation: Flange motor with Impeller/ Blower
  • Chamber Illumination: Fluorescent light with door switch
  • Feet: Castor Wheels
  • Trays: SS Wire Mesh, Heavy Duty
  • Construction: Double-walled with Poly Urethene (PUF) insulation provided
  • with outer stainless steel metal door and inner glass viewing door
walk In stability chamber manufacturer in india
Specifications
Model No. OSEPL 11
Temp. Range 10°C to 60°C
Humidity Range 40% to 95% RH
Temp. Resolution ±0.1°C ±0.1°C ±0.1°C ±0.1°C ±0.1°C ±0.1°C ±0.1°C ±0.1°C
Temp. Accuracy ±0.2°C ±0.2°C ±0.2°C ±0.2oC. ±0.2°C ±0.2°C ±0.2°C ±0.2oC.
Temp. Uniformity ±1°C ±1°C ±1°C ±1°C ±1°C ±1°C ±1°C ±1°C
Humidity Resolution 1% 1% 1% 1% 1% 1% 1% 1%
Humidity Accuracy ±2% ±2% ±2% ±2% ±2% ±2% ±2% ±2%
Humidity Uniformity ±3% ±3% ±3% ±3% ±3% ±3% ±3% ±3%
Internal Size (W x D x H) cm 2000x950x 2000 2000x1250x 2000 2000x2000x 2000 2000x2000x 2500 2000x2500x 2500 2500x2500x 2400 3000 x 3000 x 2000 3000 x 3000 x 2500
External Size (W x D x H) cm 2160 x 1360 x 2160 2160 x 1510 x 2160 2160 x 2360 x 2160 2160 x2460 x 2660 2160 x 2960 x 2660 2660 x 2960 x 2560 3120 x 3420 x 2120 3160 x 3460 x 2660
Capacity (Litres) 4000 5000 8000 10000 12500 15000 18000 22500
No. of Trays (S.S.) 12 16 24 32 36 48 5 5
No. of Racks (S.S.) 2 4 6 4 6 6 6 6
Input Voltage 230 Volts AC, 50 Hz
Material Inner Chamber: SS 304 Grade
  Outer Chamber: SS 304 Grade

 

Pre-Installation Mandatory Requirement
  1. Continuous supply of Distilled water or DM water with a tap for boiler input of ½” BSP at 3 feet height
  2. Room temperature around machine preferably at 25°C with air conditioning or a well-ventilated room with an exhaust fan. The surrounding temperature should not exceed 30°C
  3. Stabilized input voltage of 230V AC, 20 Amps. Use of Servo Controlled Stability Chamber is recommended
  4. Water drain line with ½” nozzle to be provided at floor level.
Safety Features
  • High temperature safety cut-off
  • Low water level boiler cut-off
  • Electrical overload cutoff relay for compressor
  • Time delay for compressor switch ON
Calibration / Validation
i) Calibration
  • Temperature sensor probe calibration with traceability to ERTL
  • Temperature controller calibration with traceability to ERTL
ii) Validation
  • IQ, OQ and PQ documentation and protocols shall be provided with equipment.
  • Documentation prepared / developed for Osworld by Premier Validation Ltd. – Ireland, one of Europe’s leading validation consultants
  • Performance Qualification with Temperature Mapping on site. The Performance Validation Test consists of one cycle at any one temperature point for an 8-hour period. Data is collected in printed format from up to six different points within the chamber. Temperature uniformity is considered acceptable if the deviation is less than ±1°C of set temperature.
Optional Extras / Accessories
  • Oslog Data Storage Device and Oslog 21 CFR Part 11 Compliant Software - Both have been developed exclusively for Osworld by M/s Godrej & Boyce Mfg. Co. Ltd.
  • Additional high-temperature safety system to cut off all heaters with audio-visual alarm
  • PLC-based system: PLC with HMI 3.7” touch screen grey monitor with 21 CFR Part 11 features (OSLOG) software.
  • Door access security system
  • Voltage Stabilizer for above unit
  • GSM System: Alarms sent to predefined mobile numbers
  • HMI: 5.7" touch screen grey in lieu of 3.7"
  • Controller with printer interface to connect to EPSON Dot Matrix line printer. Date, time and temperature can be printed whereas print interval can be set between 1 to 99 minutes
  • Data Logger: 4 point temp. Data logger with sensors placed at specific points in chamber with printer interface and memory